TEST AND CALIBRATION OF THE DETECTORS OF THE BES LUMINOSITY MONITOR
- Received Date:1900-01-01
- Accepted Date:1900-01-01
- Available Online:1989-10-05
Abstract:The test and calibration of the detectors of the BES luminosity monitor were performed at KEK Japan.In the momentum region from 0.25GeV/c to 4.0GeV/c,the properties of the shower counter,such as energy resolution,energy linearity,homogeneity and time resolution were measured.The energy resolution of the shower counter σEis approximately 13% at 1.5GeV/c.The time resolution is 660ps.The performance of the defining counter and the complementary counter was also measured,and the energy resolution was 30—40%,the time resolution was about 300ps.The experimental results show that the properties of all these detectors are satisfactory for the BES luminosity monitor.