Preparation of182Hf AMS Standard Sample and Test on HI-13 Tandem AMS
- Received Date:2006-11-28
- Accepted Date:2007-03-19
- Available Online:2007-10-05
Abstract:The preparation and accelerator mass spectrometry (AMS) measurements of a series of182Hf AMS calibration standards are described in this paper. The preparation consists of five processes, the synthesis of the initial182Hf/180Hf standard sample by double neutron capture reaction of180Hf, the accurate measurement of this standard for isotopic ratio with thermal ionization mass spectrometry (TIMS), the chemical separation of isobaric interference182W, the sequential dilution of the initial standard, and the preparation of HfF4. The value of182Hf/180Hf in the initial182Hf/180Hf standard was obtained with TIMS to be (1.63±0.01)×10-6. Chemical separation and purification procedures were developed and applied in the preparation of all182Hf/180Hf standards for the removal of major isobaric interference nuclide182W. As a result, three182Hf AMS standards with182Hf/180Hf ratios of (1.63±0.01)×10-6, (3.22±0.26)×10-8, (3.45±0.32)×10-10were obtained. AMS measurements of these standard samples were performed and good linearity was obtained in a plot of AMS results vs nominal values for182Hf/180Hf ranging from 1.63×10-6to 3.45×10-10.