Performance Evaluation of the HAMAMATSU SHR22000 Whole-Body PET Scanner Using the IEC Standard
- Received Date:2006-01-24
- Accepted Date:2006-02-16
- Available Online:2006-11-05
Abstract:SHR22000 is a whole-body PET scanner developed by Hamamatsu Photonics K.K. The scanner has a 600mm patient aperture and a 225mm axial FOV, which can cover the whole human head and large portion of the body so that a common whole-body scan can be accomplished within 5 bed position with a 3.2mm average transverse resolution at 10mm from the axis of FOV. The paper gives a system performance evaluation of this scanner using GB/T 18988.1-2003/IEC 61675-1:1998 standard, which includes the spatial resolution, tomographic sensitivity, count rate characteristic, scatter fraction and recovery coefficient. At last a human body scanning result is given as an imaging example.