X
Advanced Search

Experimental study of channel electron multiplier

  • BackgroundChannel electron multiplier (CEM) can be used to measure extremely few charged particles, such as electrons and ions. The CEM is widely used in particle detection, so it is very important to study their performance parameters.
    PurposesTest and analyze the performance parameters of the CEM, such as resistance, gain, and pulse output for a single entrance photoelectron.
    MethodsThe heated tantalum filament is used as a stable and adjustable planar electron source to test the performance of the CEM in the analog mode. The performance parameters of the CEM in the pulse counting mode are tested by using the ultraviolet LED to excite the gold photocathode to generate a single photoelectron.
    Results and conclusionsThe gain of the CEM in the analog mode can reach more than 10 6and the gain in the pulse counting mode can be two orders of magnitude higher. The curved helical channel has a greater advantage than the ordinary straight channel, which is conducive to weakening the ion feedback phenomenon.
  • loading
  • Binting Zhang, Cuiping Li, Shulin Liu, et al. Experimental study of channel electron multiplierJ. Radiation Detection Technology and Methods, 2023, 7(1): 117-123. DOI: 10.1007/s41605-022-00369-7
    Citation: Binting Zhang, Cuiping Li, Shulin Liu, et al. Experimental study of channel electron multiplierJ. Radiation Detection Technology and Methods, 2023, 7(1): 117-123. DOI: 10.1007/s41605-022-00369-7

Catalog

    /

      Return
      Return
        Baidu
        map