Characteristics and time resolutions of two CeBr3gamma-ray spectrometers

  • PurposeThe purpose of this work is to study the pulse shape, energy resolution, non-proportionality of energy response for gamma-rays, and time characteristics of CeBr 3detector.
    MethodThe time and energy responses of two CeBr 3detectors were characterized by an analog method and a digital method using a set of standard γ-ray sources. The pulse shape, energy resolution and non-proportionality (nPR) of energy response for γ-rays were characterized using the analog method. For the analog method, the high voltage applied to PMTs, parameters of walk and external delay of constant fraction discriminator were optimized. For the digital method, a CEAN 1729A digitizer with sampling frequency of 2 GS/s and resolution of 11 bit, and a digital constant fraction discrimination technique were used to study the time performance of the two CeBr 3detectors. Then, the coincidence time resolutions of the CeBr 3detectors for the gamma peaks of 22Na and 60Co were measured using the two methods.
    ResultsThe t rise, lifetime τ, t fallfor CeBr 321 #and 22 #are 11.2 ns, 23.8 ns, 50.2 ns, and 10.4 ns, 26.5 ns, 58.6 ns, respectively. The measured non-proportionality of CeBr 321 #and 22 #are 1.08% and 2.22%, respectively. The time resolutions of the two CeBr 3detectors are 244 ±2 ps and 248 ±3 ps at the energy peaks of 60Co source, and 336 ±2 ps and 335 ±3 ps at 511 keV for the analog and the digital methods.
    ConclusionsThe time resolutions obtained by the analog method and the digital method are almost identical. The CeBr 3detector is a good option in the applications such as half-life measurements, ToF-PET and high counting rate conditions. Furthermore, it is a good γ-ray spectrometer owing to the preferable energy resolution and non-internal activity.
  • loading
  • Jianguo Qin, Caifeng Lai, Jun Xiao, et al. Characteristics and time resolutions of two CeBr3 gamma-ray spectrometersJ. Radiation Detection Technology and Methods, 2020, 4(3): 327-336. DOI: 10.1007/s41605-020-00187-9
    Citation: Jianguo Qin, Caifeng Lai, Jun Xiao, et al. Characteristics and time resolutions of two CeBr3 gamma-ray spectrometersJ. Radiation Detection Technology and Methods, 2020, 4(3): 327-336. DOI: 10.1007/s41605-020-00187-9

Catalog

    /

      Return
      Return
        Baidu
        map